J.Strobel MD(Jena, Germany),
[273]: data by Dr. Karin Bz, Hamburg, Germany
The AcrySof IOL (model SA60AT) is an ultraviolet-absorbing . [219]: data by Dr.Hans Klett, Herne, Germany
Bausch&Lomb Akreos AO MI60 A=118.4 a0=1.49 a1=0.40 a2=0.10 pACD=5.67 sf=1.90 A=119.1 A=119.4 102 [23],[228]
[327]: data by Mr. Say Aun Quah and Mrs. C.Gore, Manchester, UK
[23]: data by M.Tomalla, MD, Duisburg, Germany
Fibronectin helps bind the material to the capsule, anchoring the lens in place to help reduce PCO and improve stability. A-constant The constant is a theoretical value that relates the lens power to AL and keratometry, it is not expressed in units and is specific to the design of the IOL and its intended location and orientation within the eye. [71]: data by Dr.S.Schmickler, Ahaus, Germany
[100]: data by K.Bump, New York, NY, USA
[113]: data by R.Arleo, MD, Ithaca, Ithaca, NY, USA
Oculentis L-302-1 A=118.0 a0= 1.833 a1=0.138 a2=0.096 pACD=5.11 sf=1.35 A=118.3 A=118.5 530 [117],[106],[141],[118]
[282]: data by Dr. Hendrik Hasche, Rostock, Germany
Oculentis L-302 discontinued A=118.5 a0= 1.30 a1=0.40 a2=0.10 pACD=5.45 sf=1.70 A=118.9 A=119.2 195 [51],[87],[105],[116],[117],[118]
(***): results based on predominantly myopic patients (mean axial lengths: 24.85 +/- 1.39 mm
These lens constants are offered only as a recommendation for a starting . [6]: based on data by different surgeons from Eye Clinic Kln-Merheim, Germany
Polytech Polylens AS66/AS66-Y A=118.2 a0= 0.939 a1=0.250 a2=0.133 pACD=5.39 sf=1.60 A=118.6 A=118.9 283 [273],[282],[287]
Oculentis Lentis Mplus LS-312MF discont A=118.0 a0= 1.03 a1=0.40 a2=0.10 pACD=5.27 sf=1.50 A=118.5 A=118.7 187 [136],[92],[259],[120],[261],[262]
SIFI Mini Well A=118.6 a0=1.33 a1=0.40 a2=0.10 pACD=5.52 sf=1.73 A=118.8 A=119.1 49 [360]
2018;12:2031-2037. [141]: data by Dr.S.Zabel, Magdeburg, Germany
[362]: data by Dipl.med.Fischer, Dr.Henjes, Schwedt, Torgau, Germany
ZEISS CT ASPHINA 404 (Acri.Lyc 44LC) A=118.0 a0=0.126 a1=0.025 a2=0.190 pACD=5.25 sf=1.49 A=118.5 A=118.7 237 [216]
[143]: data by Dr.K.Petermeier, Tbingen, Germany
[97]: data by Asim Piracha, MD, West Mifflin, PA, USA
[155]: data by Prof.Dr.Grabner, Salzburg, Austria
It therefore has a 1:1 ratio with the formulas recommended IOL power outcome. [77]: data by Dr.Baha Toygar, Istanbul, Turkey
[37]: data by F.C.Nickel, MD, Peine, Germany
[Comparison of postoperative results after implantation of hydrophilic acrylic or hydrophobic acrylic intraocular lens: data of one-year prospective clinical study]. Physicians considering lens implantation under any of the following circumstances should weigh the potential risk/benefit ratio: Patients in whom the posterior capsule is ruptured, zonules are damaged, or primary posterior capsulotomy is planned. [346]: data by Dr. Rudolf Sieben, Aachen, Germany
eyeonics AT-50SE crystalens A=119.0 a0= 0.318 a1=0.139 a2=0.182 pACD=5.62 sf=1.85 A=119.1 A=119.4 224 [115]
[375]: data by Dr. J. Juranek, Mhldorf am Inn, Deutschland
[99]: data by IOLtech
Jan 2002
[269]: data by Augenzentrum Maus, Kln, Germany
[262]: data by Dr. J.Novak, Pardurbice, Czech Republic
[263]: data by Dr. J.-M. Bosq, Nantes, France
Hoya AF-1 YA-65BB(***) (+) A=118.3 a0=1.31 a1=0.40 a2=0.10 pACD=5.65 sf=1.74 A=118.7 A=118.2 88 [95]
[404]: data by Dres. J.Brndle MD(Fssen, Germany),
[133]: data by Keith Kellum, MD, Houma, LA, USA
HumanOptics Aspira-aQA A=118.7 a0=1.35 a1=0.40 a2=0.10 pACD=5.53 sf=1.77 A=118.9 A=119.3 113 [398]
Ophtec PC265 A=118.0 a0=1.32 a1=0.40 a2=0.10 pACD=5.44 sf=1.72 A=118.9 A=119.5 63 [22]
[119]: data by Prof.Dr.R.Welt, Ludwigshafen, Germany
[68]: data by Dr.R.Walter-Fincke, Aschaffenburg, Germany
[286]: data by Dr. Martin Drner, Bocholt, Germany
If so, do not buy . Alcon AcrySof MA60AC A=118.4 a0= 0.229 a1=0.011 a2=0.205 pACD=5.67 sf=1.90 A=119.2 A=119.5 399 [18],[114],[150],[171]
[89]: data by Dr.T.Grasbon, Ingolstadt, Germany
When the mouse button is released, from Save and be link. [201]: data by Dr.J.Schock, Erlangen, Germany
Oculentis LU-313 MF30T Japan A=118.0 a0= 1.07 a1=0.40 a2=0.10 pACD=5.39 sf=1.45 A=118.1 A=117.4 37 [321]
Oshika T, Fujita Y, Inamura M, Miyata K. Mid-term and long-term clinical assessments of a new 1-piece hydrophobic acrylic IOL with hydroxyethylmeth-acrylate. Biotech Vision Eyecryl Toric A=118.5 a0=1.30 a1=0.40 a2=0.10 pACD=5.45 sf=1.66 A=118.7 A=118.9 50 [369]
[97]: data by Asim Piracha, MD, West Mifflin, PA, USA
[317]: data by Dr.K.Seftel, Halberstadt, Germany
[386]: data by Dr. CHO Beom Jin, Incheon, South Korea
[127]: data by H.Geggle MD,Seattle, WA, USA
Hoya UV55SB (Japan)(**) A=117.6 a0=1.38 a1=0.40 a2=0.10 pACD=5.57 sf=1.68 A=118.6 A=118.5 109 [80]
Constants for IOLs characterized by '(Japan)'are based on results for Japanese eyes, which tend to be somewhat longer than Caucasian eyes. [392]: data by Dr. M. Gamringer-Kroher, Weiden, Deutschland
(**): results based on predominantly myopic patients (mean axial length: 24.48 +/- 1.87 mm)
[243]: data by Jorge Buonsanti MD / Mariana Fuentes MD, Buenos Aires, Argentina
[257]: data by Dr. Klaus Mller, Innsbruck, Austria
We believe that effective technology can bring big improvement to healthcare management and patient management. C.Grenauer-Kloevekorn, Halle, Germany
[353]: data by Dr. Paul Harton, Dr. Bob Harbin, Rome, GA, USA
[202]: data by M.Franke, Kirchberg, Germany
[303]: data by Kevin H. Cuevas, MD, Golden, CO, USA
Similarly, the Design Factor of Tecnis ZCB00 is 4. What is Alcon SA60AT lens? Setting: Ecuador Intraocular lens exchanges with single-piece IOLs may be accomplished with less difficulty early rather than late. Dr.Schmidt MC6125AS A=118.4 a0=0.885 a1=0.312 a2=0.125 pACD=5.36 sf=1.60 A=118.7 A=119.1 204 [70],[122]
Physiol PodEye A=118.9 a0=1.70 a1=0.40 a2=0.10 pACD=5.85 sf=2.06 A=119.4 A=119.8 71 [351]
[72]: data by Dr.W.Htz, Bad Hersfeld, Germany
The lens has a biconvex optic with supporting haptics. Jan 2002
The Alcon Monofocal Toric IOL aspheric design delivers a broad refractive landing zone, giving you predictable distance at 20/20 and consistent functional intermediate (20/32 at 66cm) visual acuity. Medicontur Z-Flex 690 A=118.0 a0= 0.237 a1=0.315 a2=0.138 pACD=5.05 sf=1.28 A=118.2 A=118.3 437 [281]
Fritz Ruck HA-4-Q A=118.1 a0=1.14 a1=0.40 a2=0.10 pACD=5.29 sf=1.52 A=118.6 A=118.9 88 [102]
[240]: data by Dr. Heiko Lgel, Sulzbach, Germany
[253]: data by Prof.Dr.M.Bhnke, Hamburg, Germany
310 A=118.0 a0=1.01 a1=0.40 a2=0.10 pACD=5.23 sf=1.47 A=118.5 A=118.8 83 [18]
[52]: data by PD Dr.W.Sekundo, Marburg, Germany
[312]: data by Dr.Frhlich, Freiburg i.Br
Each IOL is indicated for visual correction of aphakia in adult patients following cataract surgery IQ Monofocal IOL delivers an advanced optical profile that was designed to emulate the natural crystalline lens.. Polytech Polylens A61 (+) A=118.2 a0=0.86 a1=0.40 a2=0.10 pACD=5.06 sf=1.29 A=118.2 A=118.4 79 [13],[29]
[65]: data by Dr.M.V.Partyngl, Wil, Switzerland
[331]: data by David Climenhaga MD, Edmonton, Alberta, Canada
Enhanced quality of vision. Optic glistenings were observed in 11 (11%) of 100 SA30AL cases at a mean follow-up of 8 months. [316]: data by Prof.Dr.E.Fabian, Rosenheim, Germany
[348]: data by Dr.Budo, Belgium; Dr.Krantz-Kret, The Netherlands
[133]: data by Keith Kellum, MD, Houma, LA, USA
The optical design of the To report the clinical performance characteristics of single-piece acrylic Alcon SA30AL and SA60AT intraocular lenses (IOLs). Based on 2 large datasets (1: N = 888 Hoya Vivinex aberration-correcting and 2: N = 822 Alcon SA60AT spherical lens) a classical formula constant optimisation has been performed for the Hoffer Q, Holladay 1, Haigis and Castrop formulae, to minimise the root mean squared (rms) PE (situation A). [297]: data by Carl Zeiss Meditec AG
Lens Implant SN60AT Alcon Laboratories, Inc. Oculentis LU-800 RZI A=116.0 a0= -0.11 a1=0.40 a2=0.10 pACD=4.10 sf=0.31 A=116.6 A=116.6 47 [108]
Bausch&Lomb S 122UV A=115.8 a0= -1.13 a1=0.40 a2=0.10 pACD=3.13 sf= -0.65 A=115.1 A=114.9 29 [299]
[333]: data by Dr. Alfredo Ferrer, Denia, Spain
[385]: data by Dr. GO Byung Woo, Seoul, South Korea
[224]: data by Dr.F.Bauer, Rosenheim, Germany
[329]: data by AMO Germany GmbH
[134]: data by Dr.Chr.Chassain, Montpellier, France
[349]: data by Dr.Dhaval S. Modi, Ahmedabad, India
[8]: based on data by L.Fry, MD, Garden City, KS, USA
[158]: data by Dr.Peter Fox, Chichester, UK
[231]: data by Dr.A.Bach, Schwerin, Germany
Please use at your own discretion. 1stQ Basis Q A=118.0 a0=0.278 a1=0.427 a2=0.120 pACD=5.02 sf=1.25 A=118.1 A=118.3 601 [253],[272],[245]
A retrospective review of a subgroup of 140 consecutive cases was done to categorize postoperative optic clarity.
[277]: data by Dr.M.Winter, Bremen, Germany
Staar KS-3 A=119.0 a0=1.51 a1=0.40 a2=0.10 pACD=5.67 sf=1.91 A=119.1 A=119.8 63 [56]
[218]: data by Dr.Jens Klett, Herne, Germany
All lens power below 6 diopter are available in meniscus design on the Acrysof platform (MA60MA). [46]: data by Dr.E.Welzl-Hinterkrner, Zrich, Switzerland
Ideally, outcomes should be carefully monitored and lens constants adjusted as needed (up or down) by no more than 0.1 every 10 cases. (*): A-const taken from: Mohr A, Hengerer F, Eckardt C: Retropupillare Fixation der Irisklauenlinse bei Aphakie. [106]: data by Dr.E.Becker, Oranienburg, Germany